:WMEMory<r> Commands 36
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 1169
:WMEMory<r>:LABel
(see page 1276)
Command Syntax
:WMEMory<r>:LABel <string>
<r> ::= 1 to (# ref waveforms) in NR1 format
<string> ::= quoted ASCII string
The :WMEMory<r>:LABel command sets the reference waveform label to the string
that follows.
Setting a label for a reference waveform also adds the name to the label list in
non-volatile memory (replacing the oldest label in the list).
Query Syntax
:WMEMory<r>:LABel?
The :WMEMory<r>:LABel? query returns the label associated with a particular
reference waveform.
Return Format
<string><NL>
<string> ::= quoted ASCII string
See Also • Chapter 36, “:WMEMory<r> Commands,” starting on page 1165
• ":WMEMory<r>:DISPlay" on page 1168
Label strings are 10 characters or less, and may contain any commonly used ASCII characters.
Labels with more than 10 characters are truncated to 10 characters. Lower case characters
are converted to upper case.