EasyManua.ls Logo

Keysight Technologies InfiniiVision 3000T X Series - :SBUS<n>:A429:SOURce

Keysight Technologies InfiniiVision 3000T X Series
1420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
:SBUS<n> Commands 29
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 711
:SBUS<n>:A429:SOURce
(see page 1276)
Command Syntax
:SBUS<n>:A429:SOURce <source>
<source> ::= {CHANnel<n>}
<n> ::= 1 to (# analog channels) in NR1 format
The :SBUS<n>:A429:SOURce command sets the source of the ARINC 429 signal.
Query Syntax
:SBUS<n>:A429:SOURce?
The :SBUS<n>:A429:SOURce? query returns the currently set source of the ARINC
429 signal.
Use the :TRIGger:LEVel:HIGH and :TRIGger:LEVel:LOW commands to set the
thresold levels for the selected source.
Return Format
<source><NL>
See Also ":TRIGger:LEVel:HIGH" on page 1014
":TRIGger:LEVel:LOW" on page 1015
":TRIGger:MODE" on page 1016
":SBUS<n>:MODE" on page 701
":SBUS<n>:A429:TRIGger:TYPE" on page 718
":SBUS<n>:A429:SIGNal" on page 710
":SBUS<n>:A429:SPEed" on page 712
":SBUS<n>:A429:FORMat" on page 709
"Introduction to :TRIGger Commands" on page 1007

Table of Contents

Related product manuals