:WGEN<w> Commands 35
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 1157
:WGEN<w>:OUTPut:LOAD
(see page 1276)
Command Syntax
:WGEN<w>:OUTPut:LOAD <impedance>
<w> ::= 1 or 2 in NR1 format
<impedance> ::= {ONEMeg | FIFTy}
The :WGEN<w>:OUTPut:LOAD command selects the expected output load
impedance.
The output impedance of the Gen Out BNC is fixed at 50 ohms. However, the
output load selection lets the waveform generator display the correct amplitude
and offset levels for the expected output load.
If the actual load impedance is different than the selected value, the displayed
amplitude and offset levels will be incorrect.
Query Syntax
:WGEN<w>:OUTPut:LOAD?
The :WGEN<w>:OUTPut:LOAD? query returns the current expected output load
impedance.
Return Format
<impedance><NL>
<impedance> ::= {ONEM | FIFT}
See Also "Introduction to :WGEN<w> Commands" on page 1128