:MEASure Commands 22
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 523
:MEASure:XMIN
(see page 1276)
Command Syntax
:MEASure:XMIN [<source>]
<source> ::= {CHANnel<n> | FUNCtion<m> | MATH<m> | WMEMory<r>}
<n> ::= 1-2 or 1-4 (# of analog channels) in NR1 format
<m> ::= 1 to (# math functions) in NR1 format
<r> ::= 1 to (# ref waveforms) in NR1 format
The :MEASure:XMIN command installs a screen measurement and starts an
X-at-Min-Y measurement on the selected window. If the optional source
parameter is specified, the current source is modified.
Query Syntax
:MEASure:XMIN? [<source>]
The :MEASure:XMIN? query measures and returns the horizontal axis value at
which the minimum vertical value occurs. If the optional source is specified, the
current source is modified. If all channels are off, the query returns 9.9E+37.
Return Format
<value><NL>
<value> ::= horizontal value of the minimum in NR3 format
See Also • "Introduction to :MEASure Commands" on page 458
• ":MEASure:XMAX" on page 522
• ":MEASure:TMIN" on page 1207
NOTE
:MEASure:XMIN is an alias for :MEASure:TMIN.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals