EasyManua.ls Logo

Keysight Technologies InfiniiVision 3000T X Series - :TRIGger:ZONE:STATe

Keysight Technologies InfiniiVision 3000T X Series
1420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1084 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
33 :TRIGger Commands
:TRIGger:ZONE:STATe
(see page 1276)
Command Syntax
:TRIGger:ZONE:STATe <on_off>
<on_off> ::= {{0 | OFF} | {1 | ON}}
The :TRIGger:ZONE:STATe command enables or disables the zone qualified trigger
feature.
When the zone qualified trigger is on, the zone(s) are actively being used to qualify
the trigger.
Note that the :TRIGger:ZONE<n>:STATe setting must also be ON for a zone to be
active.
Note that :TRIGger:ZONE:STATe mimics the behavior of the [Zone] key on the front
panel, and :TRIGger:ZONE<n>:STATe mimics the behavior of the Zone 1 On and
Zone 2 On softkeys. At least one zone's state must be on for the Zone Trigger
feature (:TRIGger:ZONE:STATe) to be on. When the states of both individual zones
are turned off, Zone Trigger is automatically turned off. In this case, when Zone
Trigger is turned back on Zone 1 is forced to on. Otherwise, if at least one zone
was on when Zone Trigger was turned off, the same configuration of individual
zone on/off states will be restored when Zone Trigger is turned back on.
Query Syntax
:TRIGger:ZONE:STATe?
The :TRIGger:ZONE:STATe? query returns whether the zone qualified trigger
feature is enabled or disabled.
Return Format
<on_off><NL>
<on_off> ::= {0 | 1}
See Also ":TRIGger:ZONE<n>:STATe" on page 1088
":TRIGger:ZONE:SOURce" on page 1083