1206 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
37 Obsolete and Discontinued Commands
:MEASure:TMAX
(see page 1276)
Command Syntax
:MEASure:TMAX [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= 1 to (# analog channels) in NR1 format
The :MEASure:TMAX command installs a screen measurement and starts an
X-at-Max-Y measurement on the selected waveform. If the optional source is
specified, the current source is modified.
Query Syntax
:MEASure:TMAX? [<source>]
The :MEASure:TMAX? query returns the horizontal axis value at which the
maximum vertical value occurs on the current source. If the optional source is
specified, the current source is modified. If all channels are off, the query returns
9.9E+37.
Return Format
<value><NL>
<value> ::= time at maximum in NR3 format
See Also • "Introduction to :MEASure Commands" on page 458
• ":MEASure:TMIN" on page 1207
• ":MEASure:XMAX" on page 522
• ":MEASure:XMIN" on page 523
NOTE
The :MEASure:TMAX command is obsolete and is provided for backward compatibility to
previous oscilloscopes. Use the :MEASure:XMAX command (see page 522) instead.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals