600 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
26 :POWer Commands
:POWer:HARMonics:DISPlay
(see page 1276)
Command Syntax
:POWer:HARMonics:DISPlay <display>
<display> ::= {TABLe | BAR | OFF}
The :POWer:HARMonics:DISPlay command specifies how to display the current
harmonics analysis results:
•TABLe
•BAR Bar chart.
OFF — Harmonics measurement results are not displayed.
Query Syntax
:POWer:HARMonics:DISPlay?
The :POWer:HARMonics:DISPlay query returns the display setting.
Return Format
<display><NL>
<display> ::= {TABL | BAR | OFF}
See Also ":POWer:HARMonics:APPLy" on page 598
":POWer:HARMonics:DATA" on page 599
":POWer:HARMonics:FAILcount" on page 601
":POWer:HARMonics:LINE" on page 602
":POWer:HARMonics:POWerfactor" on page 603
":POWer:HARMonics:RUNCount" on page 604
":POWer:HARMonics:STANdard" on page 605
":POWer:HARMonics:STATus" on page 606
":POWer:HARMonics:THD" on page 607