Keysight InfiniiVision 3000T X-Series Oscilloscopes
Programmer's Guide
22 :MEASure Commands
Select automatic measurements to be made and control time markers. See
"Introduction to :MEASure Commands" on page 458.
Table 100 :MEASure Commands Summary
Command Query Options and Query Returns
:MEASure:ALL (see
page 460)
n/a n/a
:MEASure:AREa
[<interval>][,<source
>] (see page 461)
:MEASure:AREa?
[<interval>][,<source
>] (see page 461)
<interval> ::= {CYCLe | DISPlay}
<source> ::= {CHANnel<n> |
FUNCtion<m> | MATH<m> |
WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<m> ::= 1 to (# math functions)
in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<return_value> ::= area in
volt-seconds, NR3 format
:MEASure:BRATe
[<source>] (see
page 462)
:MEASure:BRATe?
[<source>] (see
page 462)
<source> ::= {<digital channels>
| CHANnel<n> | FUNCtion<m> |
MATH<m> | WMEMory<r>}
<digital channels> ::=
DIGital<d> for the MSO models
<d> ::= 0 to (# digital channels
- 1) in NR1 format
<n> ::= 1 to (# of analog
channels) in NR1 format
<m> ::= 1 to (# math functions)
in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<return_value> ::= bit rate in
Hz, NR3 format