:SYSTem Commands 31
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 991
:SYSTem:SETup
(see page 1276)
Command Syntax
:SYSTem:SETup <setup_data>
<setup_data> ::= binary block data in IEEE 488.2 # format.
The :SYSTem:SETup command sets the oscilloscope as defined by the data in the
setup (learn) string sent from the controller. The setup string does not change the
interface mode or interface address.
Query Syntax
:SYSTem:SETup?
The :SYSTem:SETup? query operates the same as the *LRN? query. It outputs the
current oscilloscope setup in the form of a learn string to the controller. The setup
(learn) string is sent and received as a binary block of data. The format for the data
transmission is the # format defined in the IEEE 488.2 specification.
Return Format
<setup_data><NL>
<setup_data> ::= binary block data in IEEE 488.2 # format
See Also • "Introduction to :SYSTem Commands" on page 980
• "*LRN (Learn Device Setup)" on page 175
Example Code
' SAVE_SYSTEM_SETUP - The :SYSTEM:SETUP? query returns a program
' message that contains the current state of the instrument. Its
' format is a definite-length binary block, for example,
' #800075595<setup string><NL>
' where the setup string is 75595 bytes in length.
myScope.WriteString ":SYSTEM:SETUP?"
varQueryResult = myScope.ReadIEEEBlock(BinaryType_UI1)
CheckForInstrumentErrors ' After reading query results.
' Output setup string to a file:
Dim strPath As String
strPath = "c:\scope\config\setup.dat"
' Open file for output.
Close #1 ' If #1 is open, close it.
Open strPath For Binary Access Write Lock Write As #1
Put #1, , varQueryResult ' Write data.
Close #1 ' Close file.
' RESTORE_SYSTEM_SETUP - Read the setup string from a file and
' write it back to the oscilloscope.
Dim varSetupString As Variant
strPath = "c:\scope\config\setup.dat"
' Open file for input.
Open strPath For Binary Access Read As #1
Get #1, , varSetupString ' Read data.
Close #1 ' Close file.
' Write setup string back to oscilloscope using ":SYSTEM:SETUP"

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals