742 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
29 :SBUS<n> Commands
:SBUS<n>:CAN:TRIGger:PATTern:DATA:STARt
(see page 1276)
Command Syntax
:SBUS<n>:CAN:TRIGger:PATTern:DATA:STARt <start>
<start> ::= integer between 0 and 63, in NR1 format.
The :SBUS<n>:CAN:TRIGger:PATTern:DATA:STARt command specifies the starting
byte position for CAN FD data triggers.
CAN FD frames can have up to 64 bytes of data. You can trigger on up to 8 bytes
of data. The starting byte position setting lets you trigger on data anywhere within
the frame.
Query Syntax
:SBUS<n>:CAN:TRIGger:PATTern:DATA:STARt?
The :SBUS<n>:CAN:TRIGger:PATTern:DATA:STARt? query returns the starting byte
position setting.
Return Format
<start><NL>
<start> ::= integer between 0 and 63, in NR1 format.
See Also ":SBUS<n>:CAN:TRIGger:PATTern:DATA" on page 739
":SBUS<n>:CAN:TRIGger:PATTern:DATA:LENGth" on page 741