EasyManuals Logo

Keysight Technologies InfiniiVision 3000T X Series User Manual

Keysight Technologies InfiniiVision 3000T X Series
1420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #285 background imageLoading...
Page #285 background image
:CHANnel<n> Commands 10
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 285
:CHANnel<n>:VERNier
(see page 1276)
Command Syntax
:CHANnel<n>:VERNier <vernier value>
<vernier value> ::= {{1 | ON} | {0 | OFF}
<n> ::= 1 to (# analog channels) in NR1 format
The :CHANnel<n>:VERNier command specifies whether the channel's vernier (fine
vertical adjustment) setting is ON (1) or OFF (0).
Query Syntax
:CHANnel<n>:VERNier?
The :CHANnel<n>:VERNier? query returns the current state of the channel's
vernier setting.
Return Format
<vernier value><NL>
<vernier value> ::= {0 | 1}
See Also • "Introduction to :CHANnel<n> Commands" on page 267

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals