EasyManuals Logo

Keysight Technologies InfiniiVision 3000T X Series User Manual

Keysight Technologies InfiniiVision 3000T X Series
1420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #385 background imageLoading...
Page #385 background image
:FUNCtion<m> Commands 18
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 385
:FUNCtion<m>:FREQuency:LOWPass
(see page 1276)
Command Syntax
:FUNCtion<m>:FREQuency:LOWPass <3dB_freq>
<m> ::= 1 to (# math functions) in NR1 format
<3dB_freq> ::= -3dB cutoff frequency value in NR3 format
The :FUNCtion<m>:FREQuency:LOWPass command sets the low-pass filter's
-3 dB cutoff frequency.
The low-pass filter is a 4th order Bessel-Thompson filter.
Query Syntax
:FUNCtion<m>:FREQuency:LOWPass?
The :FUNCtion<m>:FREQuency:LOWPass query returns the low-pass filter's cutoff
frequency.
Return Format
<3dB_freq><NL>
<3dB_freq> ::= -3dB cutoff frequency value in NR3 format
See Also • ":FUNCtion<m>:OPERation" on page 390

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals