:POWer Commands 26
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 633
:POWer:SIGNals:IEXPected
(see page 1276)
Command Syntax
:POWer:SIGNals:IEXPected <value>[suffix]
<value> ::= Expected current value in NR3 format
[suffix] ::= {A | mA}
The :POWer:SIGNals:IEXPected command specifies the expected inrush current
amplitude. This value is used to set the vertical scale of the channel probing
current.
Query Syntax
:POWer:SIGNals:IEXPected?
The :POWer:SIGNals:IEXPected query returns the expected inrush current setting.
Return Format
<value><NL>
<value> ::= Expected current value in NR3 format
See Also • ":POWer:INRush:APPLy" on page 608
• ":POWer:SIGNals:AUTosetup" on page 624
• ":POWer:SIGNals:OVERshoot" on page 634
• ":POWer:SIGNals:VMAXimum:INRush" on page 635
• ":POWer:SIGNals:SOURce:CURRent<i>" on page 641
• ":POWer:SIGNals:SOURce:VOLTage<i>" on page 642