:POWer Commands 26
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 647
:POWer:SWITch:IREFerence
(see page 1276)
Command Syntax
:POWer:SWITch:IREFerence <percent>
<percent> ::= percent in NR1 format
The :POWer:SWITch:IREFerence command to specify the current switching level
for the start of switching edges. The value is in percentage of the maximum switch
current.
You can adjust this value to ignore noise floors or null offset that is difficult to
eliminate in current probes.
This value specifies the threshold that is used to determine the switching edges.
Query Syntax
:POWer:SWITch:IREFerence?
The :POWer:SWITch:IREFerence query returns the current switching level percent
value.
Return Format
<percent><NL>
<percent> ::= percent in NR1 format
See Also • ":POWer:SWITch:APPLy" on page 645
• ":POWer:SWITch:CONDuction" on page 646
• ":POWer:SWITch:RDS" on page 648
• ":POWer:SWITch:VCE" on page 649
• ":POWer:SWITch:VREFerence" on page 650

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals