1208 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
37 Obsolete and Discontinued Commands
:MEASure:TSTArt
(see page 1276)
Command Syntax
:MEASure:TSTArt <value> [suffix]
<value> ::= time at the start marker in seconds
[suffix] ::= {s | ms | us | ns | ps}
The :MEASure:TSTArt command moves the start marker (X1 cursor) to the
specified time with respect to the trigger time.
Query Syntax
:MEASure:TSTArt?
The :MEASure:TSTArt? query returns the time at the start marker (X1 cursor).
Return Format
<value><NL>
<value> ::= time at the start marker in NR3 format
See Also • "Introduction to :MARKer Commands" on page 426
• "Introduction to :MEASure Commands" on page 458
• ":MARKer:X1Position" on page 430
• ":MARKer:X2Position" on page 432
• ":MARKer:XDELta" on page 434
• ":MEASure:TDELta" on page 1204
• ":MEASure:TSTOp" on page 1209
The short form of this command, TSTA, does not follow the defined Long Form to Short Form
Truncation Rules (see page 1278). The normal short form "TST" would be the same for both
TSTArt and TSTOp, so sending TST for the TSTArt command produces an error.
The :MEASure:TSTArt command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:X1Position command (see page 430) instead.