:SYSTem Commands 31
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 987
:SYSTem:PRESet
(see page 1276)
Command Syntax
:SYSTem:PRESet
The :SYSTem:PRESet command places the instrument in a known state. This is the
same as pressing the [Default Setup] key or [Save/Recall] > Default/Erase > Default
Setup on the front panel.
When you perform a default setup, some user settings (like preferences) remain
unchanged. To reset all user settings to their factory defaults, use the *RST
command.
Reset conditions are:
Acquire Menu
Mode Normal
Averaging Off
# Averages 8
Analog Channel Menu
Channel 1 On
Channel 2 Off
Volts/division 5.00 V
Offset 0.00
Coupling DC
Probe attenuation 10:1
Vernier Off
Invert Off
BW limit Off
Impedance 1 M Ohm (cannot be changed)
Units Volts
Skew 0
Cursor Menu
Source Channel 1