:DVM Commands 15
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 337
:DVM:MODE
(see page 1276)
Command Syntax
:DVM:MODE <dvm_mode>
<dvm_mode> ::= {ACRMs | DC | DCRMs}
The :DVM:MODE command sets the digital voltmeter (DVM) mode:
• ACRMs — displays the root-mean-square value of the acquired data, with the
DC component removed.
• DC — displays the DC value of the acquired data.
• DCRMs — displays the root-mean-square value of the acquired data.
Query Syntax
:DVM:MODE?
The :DVM:MODE? query returns the selected DVM mode.
Return Format
<dvm_mode><NL>
<dvm_mode> ::= {ACRM | DC | DCRM}
See Also • ":DVM:ENABle" on page 336
• ":DVM:SOURce" on page 338
• ":DVM:ARANge" on page 334
• ":DVM:CURRent" on page 335