882 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
29 :SBUS<n> Commands
:SBUS<n>:UART:TRIGger:BURSt
(see page 1276)
Command Syntax
:SBUS<n>:UART:TRIGger:BURSt <value>
<value> ::= {OFF | 1 to 4096 in NR1 format}
The :SBUS<n>:UART:TRIGger:BURSt command selects the burst value (Nth frame
after idle period) in the range 1 to 4096 or OFF, for the trigger when in UART
mode.
Query Syntax
:SBUS<n>:UART:TRIGger:BURSt?
The :SBUS<n>:UART:TRIGger:BURSt? query returns the current UART trigger
burst value.
Return Format
<value><NL>
<value> ::= {OFF | 1 to 4096 in NR1 format}
See Also "Introduction to :TRIGger Commands" on page 1007
":TRIGger:MODE" on page 1016
":SBUS<n>:UART:TRIGger:IDLE" on page 884
":SBUS<n>:UART:TRIGger:TYPE" on page 886