:SAVE Commands 28
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 679
:SAVE:MASK[:STARt]
(see page 1276)
Command Syntax
:SAVE:MASK[:STARt] [<file_spec>]
<file_spec> ::= {<internal_loc> | <file_name>}
<internal_loc> ::= 0-3; an integer in NR1 format
<file_name> ::= quoted ASCII string
The :SAVE:MASK[:STARt] command saves a mask.
See Also "Introduction to :SAVE Commands" on page 669
":SAVE:FILename" on page 672
":SAVE:PWD" on page 682
":RECall:MASK[:STARt]" on page 662
":MTESt:DATA" on page 564
NOTE
If a file extension is provided as part of a specified <file_name>, it must be ".msk".