EasyManuals Logo

Keysight Technologies InfiniiVision 3000T X Series User Manual

Keysight Technologies InfiniiVision 3000T X Series
1420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #692 background imageLoading...
Page #692 background image
692 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
28 :SAVE Commands
:SAVE:WAVeform:LENGth
(see page 1276)
Command Syntax
:SAVE:WAVeform:LENGth <length>
<length> ::= 100 to max. length; an integer in NR1 format
When the :SAVE:WAVeform:LENGth:MAX setting is OFF, the
:SAVE:WAVeform:LENGth command sets the waveform data length (that is, the
number of points saved).
When the :SAVE:WAVeform:LENGth:MAX setting is ON, the
:SAVE:WAVeform:LENGth setting has no effect.
Query Syntax
:SAVE:WAVeform:LENGth?
The :SAVE:WAVeform:LENGth? query returns the current waveform data length
setting.
Return Format
<length><NL>
<length> ::= 100 to max. length; an integer in NR1 format
See Also • "Introduction to :SAVE Commands" on page 669
• ":SAVE:WAVeform:LENGth:MAX" on page 693
• ":SAVE:WAVeform[:STARt]" on page 690
• ":WAVeform:POINts" on page 1102
• ":SAVE:WAVeform:FORMat" on page 691

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 3000T X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 3000T X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 3000T X Series
CategoryTest Equipment
LanguageEnglish