:SBUS<n> Commands 29
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 811
:SBUS<n>:M1553:AUTosetup
(see page 1276)
Command Syntax
:SBUS<n>:M1553:TRIGger:AUTosetup
The :SBUS<n>:M1553:AUTosetup command automatically sets these options for
decoding and triggering on MIL-STD-1553 signals:
• High/Low Trigger Thresholds: to a voltage value equal to ±1/3 division based
on the source channel's current V/div setting.
• Noise Reject: Off.
• Probe Attenuation: 10.0.
•Serial Decode: On.
• Trigger: the specified serial bus (n of SBUS<n>).
See Also • "Introduction to :TRIGger Commands" on page 1007
• ":TRIGger:MODE" on page 1016
• ":SBUS<n>:M1553:SOURce" on page 813