298 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
11 :COUNter Commands
:COUNter:TOTalize:SLOPe
(see page 1276)
Command Syntax
:COUNter:TOTalize:SLOPe <slope>
<slope> ::= {{NEGative | FALLing} | {POSitive | RISing}}
The :COUNter:TOTalize:SLOPe command specifies whether positive or negative
edges are counted.
Query Syntax
:COUNter:TOTalize:SLOPe?
The :COUNter:TOTalize:SLOPe? query returns the currently set slope specification.
Return Format
<slope><NL>
<slope> ::= {NEG | POS}
See Also ":COUNter:CURRent" on page 289
":COUNter:ENABle" on page 290
":COUNter:MODE" on page 291
":COUNter:NDIGits" on page 292
":COUNter:SOURce" on page 293
":COUNter:TOTalize:CLEar" on page 294
":COUNter:TOTalize:GATE:ENABle" on page 295
":COUNter:TOTalize:GATE:POLarity" on page 296
":COUNter:TOTalize:GATE:SOURce" on page 297