1050 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
33 :TRIGger Commands
You can only specify an edge on one channel. When an edge is specified, the
:TRIGger:PATTern:QUALifier does not apply.
Query Syntax
:TRIGger:PATTern?
The :TRIGger:PATTern? query returns the pattern string, edge source, and edge.
Return Format
<string>,<edge_source>,<edge><NL>
See Also • "Introduction to :TRIGger Commands" on page 1007
• ":TRIGger:PATTern:FORMat" on page 1051
• ":TRIGger:PATTern:QUALifier" on page 1054
• ":TRIGger:MODE" on page 1016
NOTE
The optional <edge_source> and <edge> parameters should be sent together or not at all. The
edge can be specified in the ASCII <string> parameter. If the edge source and edge
parameters are used, they take precedence.