108 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
4 Commands Quick Reference
:MTESt:RMODe:TIME
<seconds> (see
page 574)
:MTESt:RMODe:TIME?
(see page 574)
<seconds> ::= from 1 to 86400 in
NR3 format
:MTESt:RMODe:WAVeform
s <count> (see
page 575)
:MTESt:RMODe:WAVeform
s? (see page 575)
<count> ::= number of waveforms
in NR1 format
:MTESt:SCALe:BIND {{0
| OFF} | {1 | ON}}
(see page 576)
:MTESt:SCALe:BIND?
(see page 576)
{0 | 1}
:MTESt:SCALe:X1
<x1_value> (see
page 577)
:MTESt:SCALe:X1? (see
page 577)
<x1_value> ::= X1 value in NR3
format
:MTESt:SCALe:XDELta
<xdelta_value> (see
page 578)
:MTESt:SCALe:XDELta?
(see page 578)
<xdelta_value> ::= X delta value
in NR3 format
:MTESt:SCALe:Y1
<y1_value> (see
page 579)
:MTESt:SCALe:Y1? (see
page 579)
<y1_value> ::= Y1 value in NR3
format
:MTESt:SCALe:Y2
<y2_value> (see
page 580)
:MTESt:SCALe:Y2? (see
page 580)
<y2_value> ::= Y2 value in NR3
format
:MTESt:SOURce
<source> (see
page 581)
:MTESt:SOURce? (see
page 581)
<source> ::= {CHANnel<n> | NONE}
<n> ::= {1 | 2 | 3 | 4} for 4ch
models
<n> ::= {1 | 2} for 2ch models
n/a :MTESt:TITLe? (see
page 582)
<title> ::= a string of up to 128
ASCII characters
Table 20 :MTESt Commands Summary (continued)
Command Query Options and Query Returns
Table 21 :POD<n> Commands Summary
Command Query Options and Query Returns
:POD<n>:DISPlay {{0 |
OFF} | {1 | ON}} (see
page 585)
:POD<n>:DISPlay? (see
page 585)
{0 | 1}
<n> ::= 1-2 in NR1 format