Commands Quick Reference 4
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 151
Table 61 :TRIGger:RUNT Commands Summary
Command Query Options and Query Returns
:TRIGger:RUNT:POLarit
y <polarity> (see
page 1057)
:TRIGger:RUNT:POLarit
y? (see page 1057)
<polarity> ::= {POSitive |
NEGative | EITHer}
:TRIGger:RUNT:QUALifi
er <qualifier> (see
page 1058)
:TRIGger:RUNT:QUALifi
er? (see page 1058)
<qualifier> ::= {GREaterthan |
LESSthan | NONE}
:TRIGger:RUNT:SOURce
<source> (see
page 1059)
:TRIGger:RUNT:SOURce?
(see page 1059)
<source> ::= CHANnel<n>
<n> ::= 1 to (# analog channels)
in NR1 format
:TRIGger:RUNT:TIME
<time>[suffix] (see
page 1060)
:TRIGger:RUNT:TIME?
(see page 1060)
<time> ::= floating-point number
in NR3 format
[suffix] ::= {s | ms | us | ns |
ps}
Table 62 :TRIGger:SHOLd Commands Summary
Command Query Options and Query Returns
:TRIGger:SHOLd:SLOPe
<slope> (see
page 1062)
:TRIGger:SHOLd:SLOPe?
(see page 1062)
<slope> ::= {NEGative | POSitive}
:TRIGger:SHOLd:SOURce
:CLOCk <source> (see
page 1063)
:TRIGger:SHOLd:SOURce
:CLOCk? (see
page 1063)
<source> ::= {CHANnel<n> |
DIGital<d>}
<n> ::= 1 to (# analog channels)
in NR1 format
<d> ::= 0 to (# digital channels
- 1) in NR1 format
:TRIGger:SHOLd:SOURce
:DATA <source> (see
page 1064)
:TRIGger:SHOLd:SOURce
:DATA? (see page 1064)
<source> ::= {CHANnel<n> |
DIGital<d>}
<n> ::= 1 to (# analog channels)
in NR1 format
<d> ::= 0 to (# digital channels
- 1) in NR1 format
:TRIGger:SHOLd:TIME:H
OLD <time>[suffix]
(see page 1065)
:TRIGger:SHOLd:TIME:H
OLD? (see page 1065)
<time> ::= floating-point number
in NR3 format
[suffix] ::= {s | ms | us | ns |
ps}
:TRIGger:SHOLd:TIME:S
ETup <time>[suffix]
(see page 1066)
:TRIGger:SHOLd:TIME:S
ETup? (see page 1066)
<time> ::= floating-point number
in NR3 format
[suffix] ::= {s | ms | us | ns |
ps}