392 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
18 :FUNCtion<m> Commands
• MINHold — Operates on a single analog channel source or on a lower math
function. The Min Hold (or Min Envelope) operator records the minimum
vertical values found at each horizontal bucket across multiple analysis
cycles and uses those values to build a waveform.
• TRENd — Measurement trend — Operates on a single analog channel source.
The TRENd:MEASurement command selects the measurement whose trend
you want to measure.
• BTIMing — Chart logic bus timing — Operates on a bus made up of digital
channels. The BUS:YINcrement, BUS:YORigin, and BUS:YUNits commands
specify function values.
• BSTate — Chart logic bus state — Operates on a bus made up of digital
channels. The BUS:YINcrement, BUS:YORigin, and BUS:YUNit commands
specify function values. The BUS:CLOCk and BUS:SLOPe commands specify
the clock source and edge.
Query Syntax
:FUNCtion<m>:OPERation?
The :FUNCtion<m>:OPERation? query returns the current operation for the
selected function.
Return Format
<operation><NL>
<operation> ::= {ADD | SUBT | MULT | DIV | INT | DIFF | FFT | SQRT
| MAGN | ABS | SQU | LN | LOG | EXP | TEN | LOWP | HIGH | AVER
| SMO | ENV | LIN | MAXH | MINH | TREN | BTIM | BST}
See Also • "Introduction to :FUNCtion<m> Commands" on page 368
• ":FUNCtion<m>:SOURce1" on page 398
• ":FUNCtion<m>:SOURce2" on page 400
• ":FUNCtion<m>:INTegrate:IOFFset" on page 386
• ":FUNCtion<m>[:FFT]:SPAN" on page 381
• ":FUNCtion<m>[:FFT]:CENTer" on page 378
• ":FUNCtion<m>[:FFT]:VTYPe" on page 382
• ":FUNCtion<m>[:FFT]:WINDow" on page 383
• ":FUNCtion<m>:LINear:GAIN" on page 387
• ":FUNCtion<m>:LINear:OFFSet" on page 388
• ":FUNCtion<m>:FREQuency:LOWPass" on page 385
• ":FUNCtion<m>:FREQuency:HIGHpass" on page 384
• ":FUNCtion<m>:AVERage:COUNt" on page 370
• ":FUNCtion<m>:TRENd:MEASurement" on page 401
• ":FUNCtion<m>:BUS:YINCrement" on page 373
• ":FUNCtion<m>:BUS:YORigin" on page 374
• ":FUNCtion<m>:BUS:YUNits" on page 375