450 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
22 :MEASure Commands
:MEASure:PERiod
[<source>] (see
page 488)
:MEASure:PERiod?
[<source>] (see
page 488)
<source> ::= {CHANnel<n> |
FUNCtion<m> | MATH<m> |
WMEMory<r>} for DSO models
<source> ::= {CHANnel<n> |
DIGital<d> | FUNCtion<m> |
MATH<m> | WMEMory<r>} for MSO
models
<n> ::= 1 to (# analog channels)
in NR1 format
<m> ::= 1 to (# math functions)
in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<d> ::= 0 to (# digital channels
- 1) in NR1 format
<return_value> ::= waveform
period in seconds in NR3 format
:MEASure:PHASe
[<source1>]
[,<source2>] (see
page 489)
:MEASure:PHASe?
[<source1>]
[,<source2>] (see
page 489)
<source1,2> ::= {CHANnel<n> |
FUNCtion<m> | MATH<m> |
WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<m> ::= 1 to (# math functions)
in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<return_value> ::= the phase
angle value in degrees in NR3
format
:MEASure:PPULses
[<source>] (see
page 490)
:MEASure:PPULses?
[<source>] (see
page 490)
<source> ::= {CHANnel<n> |
FUNCtion<m> | MATH<m> |
WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<m> ::= 1 to (# math functions)
in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<return_value> ::= the rising
pulse count in NR3 format
Table 100 :MEASure Commands Summary (continued)
Command Query Options and Query Returns