EasyManua.ls Logo

Keysight Technologies InfiniiVision 3000T X Series - Page 454

Keysight Technologies InfiniiVision 3000T X Series
1420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
454 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
22 :MEASure Commands
n/a :MEASure:TVALue?
<value>,
[<slope>]<occurrence>
[,<source>] (see
page 509)
<value> ::= voltage level that
the waveform must cross.
<slope> ::= direction of the
waveform when <value> is crossed.
<occurrence> ::= transitions
reported.
<source> ::= {CHANnel<n> |
FUNCtion<m> | MATH<m> |
WMEMory<r>} for DSO models
<source> ::= {CHANnel<n> |
DIGital<d> | FUNCtion<m> |
MATH<m> | WMEMory<r>} for MSO
models
<n> ::= 1 to (# analog channels)
in NR1 format
<m> ::= 1 to (# math functions)
in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<d> ::= 0 to (# digital channels
- 1) in NR1 format
<return_value> ::= time in
seconds of specified voltage
crossing in NR3 format
:MEASure:VAMPlitude
[<source>] (see
page 511)
:MEASure:VAMPlitude?
[<source>] (see
page 511)
<source> ::= {CHANnel<n> |
FUNCtion<m> | MATH<m> |
WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<m> ::= 1 to (# math functions)
in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<return_value> ::= the amplitude
of the selected waveform in volts
in NR3 format
Table 100 :MEASure Commands Summary (continued)
Command Query Options and Query Returns

Table of Contents

Related product manuals