EasyManua.ls Logo

Keysight Technologies InfiniiVision 3000T X Series - Page 471

Keysight Technologies InfiniiVision 3000T X Series
1420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
:MEASure Commands 22
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 471
90%, 50%, and 10% values between Vbase and Vtop. If you want to move the
delay measurement point nearer to Vtop or Vbase, you must change the threshold
values with the :MEASure:DEFine THResholds command.
Return Format
<value><NL>
<value> ::= floating-point number delay time in seconds in NR3 format
See Also "Introduction to :MEASure Commands" on page 458
":MEASure:DEFine" on page 467
":MEASure:PHASe" on page 489

Table of Contents

Related product manuals