:MEASure Commands 22
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide 471
90%, 50%, and 10% values between Vbase and Vtop. If you want to move the
delay measurement point nearer to Vtop or Vbase, you must change the threshold
values with the :MEASure:DEFine THResholds command.
Return Format
<value><NL>
<value> ::= floating-point number delay time in seconds in NR3 format
See Also • "Introduction to :MEASure Commands" on page 458
• ":MEASure:DEFine" on page 467
• ":MEASure:PHASe" on page 489