EasyManua.ls Logo

Keysight Technologies InfiniiVision 3000T X Series - Page 698

Keysight Technologies InfiniiVision 3000T X Series
1420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
698 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
29 :SBUS<n> Commands
IIC (Inter-IC bus) triggering— consists of connecting the oscilloscope to the
serial data (SDA) line and the serial clock (SCL) line, then triggering on a
stop/start condition, a restart, a missing acknowledge, or on a read/write frame
with a specific device address and data value.
LIN (Local Interconnect Network) triggering— will trigger on LIN sync break at
the beginning of a message frame.You can trigger on Sync Break, Frame IDs, or
Frame IDs and Data.
SPI (Serial Peripheral Interface) triggering— consists of connecting the
oscilloscope to a clock, data (MOSI or MISO), and framing signal. You can then
trigger on a data pattern during a specific framing period. The serial data string
can be specified to be from 4 to 64 bits long.
UART/RS-232 triggering (with COMP license) — lets you trigger on RS-232
serial data.
SENT triggering (with SENSOR license) — lets you trigger on SENT serial data.
Reporting the Setup
Use :SBUS<n>? to query setup information for the :SBUS<n> subsystem.
Return Format
The following is a sample response from the :SBUS1? query. In this case, the query
was issued following a *RST command.
:SBUS1:DISP 0;MODE IIC;:SBUS1:IIC:ASIZ BIT7;:SBUS1:IIC:TRIG:TYPE
STAR;QUAL EQU;:SBUS1:IIC:SOUR:CLOC CHAN1;DATA
CHAN2;:SBUS1:IIC:TRIG:PATT:ADDR -1;DATA -1;DATA2 -1
NOTE
Two I2S buses or two SPI buses cannot be decoded on both SBUS1 and SBUS2 at the same
time.

Table of Contents