EasyManua.ls Logo

Keysight Technologies InfiniiVision 3000T X Series - Page 92

Keysight Technologies InfiniiVision 3000T X Series
1420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
92 Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
4 Commands Quick Reference
Table 19 :MEASure Commands Summary
Command Query Options and Query Returns
:MEASure:ALL (see
page 460)
n/a n/a
:MEASure:AREa
[<interval>][,<source
>] (see page 461)
:MEASure:AREa?
[<interval>][,<source
>] (see page 461)
<interval> ::= {CYCLe | DISPlay}
<source> ::= {CHANnel<n> |
FUNCtion<m> | MATH<m> |
WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<m> ::= 1 to (# math functions)
in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<return_value> ::= area in
volt-seconds, NR3 format
:MEASure:BRATe
[<source>] (see
page 462)
:MEASure:BRATe?
[<source>] (see
page 462)
<source> ::= {<digital channels>
| CHANnel<n> | FUNCtion<m> |
MATH<m> | WMEMory<r>}
<digital channels> ::=
DIGital<d> for the MSO models
<d> ::= 0 to (# digital channels
- 1) in NR1 format
<n> ::= 1 to (# of analog
channels) in NR1 format
<m> ::= 1 to (# math functions)
in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<return_value> ::= bit rate in
Hz, NR3 format
:MEASure:BWIDth
[<source>] (see
page 463)
:MEASure:BWIDth?
[<source>] (see
page 463)
<source> ::= {CHANnel<n> |
FUNCtion<m> | MATH<m> |
WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<m> ::= 1 to (# math functions)
in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<return_value> ::= burst width in
seconds, NR3 format
:MEASure:CLEar (see
page 464)
n/a n/a

Table of Contents

Related product manuals