3 Spectrum Analyzer Mode
3.8 Spurious Emissions Measurement
This parameter can send up to 20 values. The location in the list sent corresponds to
the range of the associated value. When sending the remote command, missing
values are not permitted. If you want to change values 2 and 6, then you must send
all values up to 6. Subsequent values will remain as they were.
The query always returns 20 values.
Remote
Command
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:TIME <time>, <time>, <time>, <time>,
<time>, <time>, <time>, <time>, <time>, <time>, <time>, <time>, <time>,
<time>, <time>, <time>, <time>, <time>, <time>, <time>
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:TIME?
Example
:SPUR:SWE:TIME 10,10,10,10,10,10,10,10,10,10,10,10,10,10,10,10,10,10,10,10
:SPUR:SWE:TIME?
Preset Automatically calculated
State Saved Saved in instrument state
Min/Max 1.0E-3/2.0E+3
Auto Function
Remote
Command
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:TIME:AUTO OFF | ON | 0 | 1,OFF | ON |
0 | 1,OFF | ON | 0 | 1,OFF | ON | 0 | 1,OFF | ON | 0 | 1,OFF | ON | 0 | 1,OFF
| ON | 0 | 1,OFF | ON | 0 | 1,OFF | ON | 0 | 1,OFF | ON | 0 | 1,OFF | ON | 0
| 1,OFF | ON | 0 | 1,OFF | ON | 0 | 1,OFF | ON | 0 | 1,OFF | ON | 0 | 1,OFF |
ON | 0 | 1,OFF | ON | 0 | 1,OFF | ON | 0 | 1,OFF | ON | 0 | 1,OFF | ON | 0 |
1
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:TIME:AUTO?
Example
:SPUR:SWE:TIME:AUTO
ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON
:SPUR:SWE:TIME:AUTO?
Preset
ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON,
ON
State Saved Saved in instrument state
Range
OFF|ON
Points
Sets the number of points per sweep for the measurement. This parameter can send
up to 20 values. The location in the list sent corresponds to the range of the
associated value. When sending the remote command, missing values are not
permitted. If you want to change values 2 and 6, then you must send all values up to
6. Subsequent values will remain as they were. The query always returns 20 values.
The Points mode can be manual, where you determine the setting or auto, where the
instrument determines the number of trace points to ensure the sweep points
resolution equals RBW/2. This is calculated using the following algorithm:
Points = (Stop Freq – Start Freq) / (ResBW / 2)
1462 Spectrum Analyzer Mode User's &Programmer's Reference