3 Spectrum Analyzer Mode
3.8 Spurious Emissions Measurement
This parameter can send up to 20 values. The location in the list sent corresponds to
the range of the associated value. When sending the remote command, missing
values are not permitted. If you want to change values 2 and 6, then you must send
all values up to 6. Subsequent values will remain as they were.
The query always returns 20 values.
Remote
Command
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:TIME <time>, <time>, <time>, <time>,
<time>, <time>, <time>, <time>, <time>, <time>, <time>, <time>, <time>,
<time>, <time>, <time>, <time>, <time>, <time>, <time>
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:TIME?
Example
:SPUR:SWE:TIME 10,10,10,10,10,10,10,10,10,10,10,10,10,10,10,10,10,10,10,10
:SPUR:SWE:TIME?
Preset Automatically calculated
State Saved Saved in instrument state
Min/Max 1.0E-3/2.0E+3
Auto Function
Remote
Command
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:TIME:AUTO OFF | ON | 0 | 1,OFF | ON |
0 | 1,OFF | ON | 0 | 1,OFF | ON | 0 | 1,OFF | ON | 0 | 1,OFF | ON | 0 | 1,OFF
| ON | 0 | 1,OFF | ON | 0 | 1,OFF | ON | 0 | 1,OFF | ON | 0 | 1,OFF | ON | 0
| 1,OFF | ON | 0 | 1,OFF | ON | 0 | 1,OFF | ON | 0 | 1,OFF | ON | 0 | 1,OFF |
ON | 0 | 1,OFF | ON | 0 | 1,OFF | ON | 0 | 1,OFF | ON | 0 | 1,OFF | ON | 0 |
1
[:SENSe]:SPURious[:RANGe][:LIST]:SWEep:TIME:AUTO?
Example
:SPUR:SWE:TIME:AUTO
ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON
:SPUR:SWE:TIME:AUTO?
Preset
ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON,
ON
State Saved Saved in instrument state
Range
OFF|ON
Points
Sets the number of points per sweep for the measurement. This parameter can send
up to 20 values. The location in the list sent corresponds to the range of the
associated value. When sending the remote command, missing values are not
permitted. If you want to change values 2 and 6, then you must send all values up to
6. Subsequent values will remain as they were. The query always returns 20 values.
The Points mode can be manual, where you determine the setting or auto, where the
instrument determines the number of trace points to ensure the sweep points
resolution equals RBW/2. This is calculated using the following algorithm:
Points = (Stop Freq Start Freq) / (ResBW / 2)
1462 Spectrum Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals