3 Spectrum Analyzer Mode
3.8 Spurious Emissions Measurement
| 1, OFF | ON | 0 | 1, OFF | ON | 0 | 1
[:SENSe]:SPURious[:RANGe][:LIST]:BANDwidth[:RESolution]:AUTO?
Example
:SPUR:BWID:AUTO ON, ON, ON, OFF, OFF, OFF, OFF, OFF, ON, ON, ON, ON, OFF,
OFF, OFF, OFF, OFF, OFF, ON, ON
:SPUR:BWID:AUTO?
Preset
SA Mode
OFF, OFF, OFF, OFF, OFF, ON, ON, ON, ON, ON, ON,
ON, ON, ON, ON, ON, ON, ON, ON, ON
WCDMA Mode
OFF, OFF, OFF, OFF, OFF, OFF, OFF, OFF, ON, ON,
ON, ON, ON, ON, ON, ON, ON, ON, ON, ON
LTE, LTEAFDD, LTETDD,
LTEATDD, 5G NR Modes
OFF, OFF, OFF, OFF, OFF, OFF, OFF, ON, ON, ON,
ON, ON, ON, ON, ON, ON, ON, ON, ON, ON
WLAN Mode
OFF, OFF, OFF, OFF, ON, ON, ON, ON, ON, ON, ON,
ON, ON, ON, ON, ON, ON, ON, ON, ON
MSR Mode
OFF, OFF, OFF, OFF, OFF, OFF, OFF, ON, ON, ON,
ON, ON, ON, ON, ON, ON, ON, ON, ON, ON
State Saved Saved in instrument state
Backwards
Compatibility
SCPI
[:SENSe]:SPURious[:RANGe][:LIST]:BWIDth[:RESolution]:AUTO
Meas BW
Allows you to specify a multiplier of Res BW for the measurement integration
bandwidth.
Meas BW is multiplier integer number. It shows a ratio between Integration BW and
Resolution BW of the measurement result:
Integ BW = Meas BW * Resolution BW
Integration BW is desired resolution bandwidth and Resolution BW is actual
bandwidth for sweep. Measurement sweeps with Resolution BW and Meas BW
compensates sweep resolution bandwidth to Integration BW.
When sending the remote command, missing values are not permitted; that is, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values
will remain as they were. The query always returns 20 values.
Remote
Command
[:SENSe]:SPURious[:RANGe][:LIST]:BANDwidth:IMULti <integer>, …
[:SENSe]:SPURious[:RANGe][:LIST]:BANDwidth:IMULti?
Example
:SPUR:BAND:IMUL 1,1,1,1,1,1
:SPUR:BAND:IMUL?
Spectrum Analyzer Mode User's &Programmer's Reference 1447