6 Input/Output
6.1 RF Source
6 Power
<power>
Data Type: ampl
Specifies the output power for the step in dBm. For details of the valid ranges see
"Power" on page 2248
7 Waveform
<waveform>
Data Type: string
Specifies the waveform for playback during the step. The step can output either a
new ARB waveform, continue playback of the previous waveform, or output a CW
tone. The options for specifying these are:
<filename>
Plays the specified waveform from the start. The filename
value is the name of the file within ARB playback memory,
it is does not include the windows path to the file on the
HDD. If you enter a filename for a waveform that does not
reside within ARB playback memory, an error is
generated
CONT
Continues playback of the ARB file from the previous step
CW
Outputs a CW tone
OFF
Disables RF output
For further details, see "Waveform" on page 2248 and "Waveform File" on page
2249
8 Step Duration
<duration>
Data Type: enum
Specifies the duration of the step, as one of:
TIME|COUNt|CONTinuous
The duration can be specified to be either time, or play count of the ARB file
associated with the step, or continuous. If Waveform is set to CW, this value cannot
be set to Play Count and an error will be generated. If CONTinuous is selected, the
following Time or Count value is ignored. For further details, see "Step Duration" on
page 2249
9 Time or Count
<time_count>
Data Type: time/int
Specifies time duration in seconds, or play count of the ARB file associated with the
step
For further details, see "Play Count" on page 2251
10 Output Trigger
<trig_state>
Data Type: boolean
Specifies the output trigger state for the step, as one of:
ON|OFF|1|0
For further details, see "Trig Out" on page 2252
Step Configuration of Step Trigger parameter list (Remote Command Only)
Configures the “Step Trigger” parameter array of the whole List Sequencer at one
time. The number of arrays is the same as the step number defined in "Number of
2254 Spectrum Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)