3 Spectrum Analyzer Mode
3.5 ACP Measurement
Carrier Pwr Present
Configures the carriers for this measurement. Allows spaces to be inserted between
carriers. Carriers with the power present parameter set to YES are carriers, and
those with the power present parameter set to NO are spaces. Each carrier power
present is set to YES or NO. The individual carrier can be set by selecting the desired
carrier.
The query returns the current values for all of the carriers. If a carrier is defined as
having no power present, the power displayed is relative to the reference carrier,
otherwise the absolute power is displayed.
If you change the carrier power present to NO, and that carrier is currently
configured as the reference carrier, the next carrier to the left (or the right if there
are no carriers to the left) is assigned as the reference carrier. This also applies to
the scenario where there are only two carriers configured as having power present,
and you configure only one carrier to have no power present.
Remote
Command
[:SENSe]:ACPower:CARRier[1]|2:LIST:PPResent YES | NO, …
[:SENSe]:ACPower:CARRier[1]|2:LIST:PPResent?
Subopcode: 1 = BTS/Downlink (Default), 2 = MS/Uplink
Example
:ACP:CARR2:LIST:PPR YES
:ACP:CARR2:LIST:PPR?
Notes Some Modes do not support Carrier subopcode 2. In those cases, commands with subopcode 2 are
accepted without error but ignored
When setting these values remotely, the position in the list sent corresponds to the carrier. Missing
values are not permitted, so if you want to change values 2 and 6, you must send all values up to 6.
Subsequent values remain unchanged, unless the number of values sent is greater than the number of
carriers, then subsequent values are ignored
For LTEAFDD or LTEATDD Modes, this control is not shown. To maintain backwards compatibility with
legacy LTE FDD/TDD Modes, the SCPI command is supported in the LTE & LTE-A converged applic-
ations
Couplings Coupled to number of carriers. When the SCPI command is sent, the number of carriers is set to the
number of entries in the parameter list
Preset
YES
State Saved Saved in instrument state
Backwards
Compatibility
SCPI
[:SENSe]:MCPower:CARRier[1]|2:LIST:PPResent
1016 Spectrum Analyzer Mode User's &Programmer's Reference