9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
Conducted alignment failed bit being cleared will be reported
Preset 0
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
Questionable Calibration Extended Failure Positive Transition
Determines which bits in the "Questionable Calibration Extended Failure Condition"
on page 2814 register will set the corresponding bit in the "Questionable Calibration
Extended Failure Event Query" on page 2815 register when the condition register bit
has a positive transition (0 to 1).
The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Remote
Command
:STATus:QUEStionable:CALibration:EXTended:FAILure:PTRansition <integer>
:STATus:QUEStionable:CALibration:EXTended:FAILure:PTRansition?
Example
:STAT:QUES:CAL:EXT:FAIL:PTR 1
Conducted alignment failed bit being set will be reported
Preset 32767
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
9.4.6.12 Questionable Calibration Skipped Register
Monitors conditions that occur because a calibration or alignment has been skipped
due to various settings or conditions. It summarizes them in bit 11 of the
"Questionable Calibration Register" on page 2806.
Spectrum Analyzer Mode User's &Programmer's Reference 2816