9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
Bit Condition Operation
0 Align RF
skipped
During an alignment, the calibration of the RF section (frontend) of the
instrument was not performed. This can be caused by an interfering
user signal present at the RF Input
See "Align Now" on page 2108, "Align Now All" on page 2110
Filter Registers
"Questionable Calibration Skipped Condition" on page 2817
"Questionable Calibration Skipped Enable" on page 2818
"Questionable Calibration Skipped Event Query" on page 2818
"Questionable Calibration Skipped Negative Transition" on page 2818
"Questionable Calibration Skipped Positive Transition" on page 2819
Questionable Calibration Skipped Condition
Returns the decimal value of the sum of the bits in the Questionable Calibration
Skipped Condition register.
NOTE
The data in this register is continuously updated and reflects the current
conditions.
Remote
Command
:STATus:QUEStionable:CALibration:SKIPped:CONDition?
Example
:STAT:QUES:CAL:SKIP:COND?
Preset 0
Status Bits/OPC
dependencies
Sequential command
2817 Spectrum Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals