3 Spectrum Analyzer Mode
3.8 Spurious Emissions Measurement
:POW:PADJ?
Notes The value on the control is displayed to 0.1MHz resolution
Dependencies
Does not appear in CXA-m
Does not appear in VXT Models M9410A/11A/15A/16A
Does not appear in M9410E/11E/15E/16E
Grayed-out if microwave preselector is off
Grayed-out if entirely in Band 0, that is, if Stop Freq is lower than about 3.6 GHz
Grayed-out if entirely above 50GHz, that is, if Start Freq is higher than 50GHz
Blank in models that do not include a preselector, such as Option503. If the command is sent in
these instruments, it is accepted without error, and the query always returns 0
Grayed-out in the Spectrogram View
Preset 0MHz
State Saved The Preselector Adjust value set by "Presel Center" on page 1978, or by manually adjusting
Preselector Adjust
Not saved in instrument state, and does not survive a Preset or power cycle
Min/Max –/+500MHz
Backwards
Compatibility SCPI
[:SENSe]:POWer[:RF]:MW:PADJust
[:SENSe]:POWer[:RF]:MMW:PADJust
Backwards Compatibility Command
Notes The command has no effect, and the query always returns MWAVe
Backwards
Compatibility
SCPI
[:SENSe]:POWer[:RF]:PADJust:PRESelector MWAVe | MMWave | EXTernal
[:SENSe]:POWer[:RF]:PADJust:PRESelector?
Internal Preamp
Accesses a menu of controls for the internal preamps. Turning on the preamp gives
a better noise figure, but a poorer inter-modulation distortion (TOI) to noise floor
dynamic range. You can optimize this setting for your measurement.
The instrument takes the preamp gain into account as it sweeps. If you sweep
outside of the range of the preamp, the instrument will also account for that. The
displayed result always reflects the correct gain.
For some measurements, when the preamp is on and any part of the displayed
frequency range is below the lowest frequency for which the preamp has
specifications, a warning condition message appears in the status line. For example,
Spectrum Analyzer Mode User's &Programmer's Reference 1389

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals