3 Spectrum Analyzer Mode
3.13 Phase and Amplitude vs Time Measurement
n Results Returned
For example if "Segments" on page 2013 = 5, then the return result would look like:
0, 0, 0, -168E-03, -9.63E-03, 0.9212E-03, 0.999E-03, 161.455E-03, 70.976E-03, -0.612E-03, 2.000E-03,
-87.325E-03, 59.676E-03, -0.347E-03, 2.999E-03, 301.646E-03, 104.21E-03, -0.318E-03, 4.000E-03,
358.402E-03
2 Returns scalar Phase results for the segments measured, equivalent to the Phase values from the n = 1
query
For example if Segments = 5, then the return result would look like:
0,10,-10,20,-30
3 Returns scalar Amplitude results for the segments measured, equivalent to the Amplitude values from the n
= 1 query
For example if Segments = 5, then the return result would look like:
0,-.5,-1.2,1.2,2.5
4 Returns scalar Time results for the segments measured, equivalent to the Time values from the n = 1 query
For example if Segments = 5, then the return result would look like:
0,10,20,30,40
5 Returns scalar Frequency Error results for the segments measured
For example if Segments = 5, then the return result would look like:
0,100000,120000,-80000,90000
6 Returns the Relative Baseline results
1.
Phase
2.
Amplitude
For example: 125, -15.25
3.13.1 Views
The PAvT measurement has one view: "Normal" on page 1998.
This is a multiple-window View. When in a multiple-window View, you select a
window by touching it. The menu controls may sometimes change depending on
which window is selected.
3.13.1.1 Normal
Windows: "Graph" on page 1950, "Results" on page 1950
The PAvT measurement provides tabular result and graphical display.
Spectrum Analyzer Mode User's &Programmer's Reference 1949