9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
–
"Questionable Integrity Condition" on page 2821
–
"Questionable Integrity Enable" on page 2821
–
"Questionable Integrity Event Query" on page 2822
–
"Questionable Integrity Negative Transition" on page 2822
–
"Questionable Integrity Positive Transition" on page 2822
Questionable Integrity Condition
Returns the decimal value of the sum of the bits in the Questionable Integrity
Condition register.
NOTE
The data in this register is continuously updated and reflects the current
conditions.
Remote Command
:STATus:QUEStionable:INTegrity:CONDition?
Example
:STAT:QUES:INT:COND?
Preset 0
Status Bits/OPC depend-
encies
Sequential command
Questionable Integrity Enable
Determines which bits in the "Questionable Integrity Condition" on page 2821
Register will set bits in the "Questionable Integrity Event Query" on page 2822
register, which also sets the Integrity Summary bit (bit 9) in the "Questionable
Register" on page 2794.
The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Remote
Command
:STATus:QUEStionable:INTegrity:ENABle <integer>
:STATus:QUEStionable:INTegrity:ENABle?
Example
:STAT:QUES:INT:ENAB 8
Uncalibrated Summary will be reported to the Integrity Summary of the Status Questionable register
Preset 32767
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
2821 Spectrum Analyzer Mode User's &Programmer's Reference