9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
The variable <integer> is the sum of the decimal values of the bits that you want to
enable.
Remote
Command
:STATus:QUEStionable:INTegrity:PTRansition <integer>
:STATus:QUEStionable:INTegrity:PTRansition?
Example
:STAT:QUES:INT:PTR 8
Uncalibrated Summary being set will be reported to the Integrity Summary of the Status Questionable
register
Preset 32767
Min 0
Max 32767
Status Bits/OPC
dependencies
Sequential command
9.4.6.14 Questionable Integrity Signal Register
Monitors conditions that occur because a measurement may not be able to return
an accurate or valid result due to signal conditions. It summarizes them in bit 0 of
the "Questionable Integrity Register" on page 2819.
Bit Condition Operation
2 Burst not
found
The instrument is expecting a bursted signal but such a signal cannot
be detected because of inappropriate parameter settings or incorrect
signal content
3 Timing Error The instrument cannot establish appropriate timing from the signal
4 Carrier(s)
incorrect or
missing
The instrument cannot find the expected carrier(s) within the frequency
ranges in which it is looking
5 Frequency out
of range
One or more system or signal input frequencies are out of range
6 Sync error The instrument cannot establish sync with the measured signal
7 Demod error The instrument cannot demodulate the signal due to inappropriate
2823 Spectrum Analyzer Mode User's &Programmer's Reference