EasyManuals Logo

Keysight x-series User Manual

Keysight x-series
2852 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #2237 background imageLoading...
Page #2237 background image
6 Input/Output
6.1 RF Source
Once the List Sequencer has been configured using the front panel, SCPI, or by
loading a tab-delimited file, the sequence must be initiated using the front panel
Initiate Sequence key, or the corresponding SCPI command.
Dependencies Not available in E7760B
6.1.8.1 Sequencer
Sets the state of "List Sequencer" on page 2236
When List Sequencer is ON, the source outputs the sequence defined by the
sequencer
When List Sequencer is OFF, the source outputs a single waveform segment or
sequence (independent mode) at a single frequency and amplitude
Remote Command
:SOURce:LIST[:STATe] ON | OFF | 1 | 0
:SOURce:LIST[:STATe]?
Example
:SOUR:LIST OFF
Notes When the sequencer is ON, the List Sequencer controls the output of the source
Dependencies Not available in E7760B
Couplings When in Sequence Analyzer Mode, and the List Sequencer state is OFF, Include Source is forced to
NO, and the Include Source key is grayed-out
When in Sequence Analyzer Mode, and the List Sequencer state is ON, Include Source is available to
set, and an ARB memory related operation such as load or delete will be rejected
Preset
OFF
Range
ON | OFF
6.1.8.2 Initiate Sequence
Arms the sequence for single execution. Once the sequence is armed, the source
begins the sequence as soon as the trigger is received. If trigger is set to Free Run,
the sequence starts immediately.
Remote Command
:SOURce:LIST:TRIGger[:IMMediate]
Example
:SOUR:LIST:TRIG
Notes When in Sequence Analyzer Mode, and Include Source is ON, the Initiate List Sequencer operation is
rejected, and the key is grayed-outIf the file needed by the sequencer is not already in ARB memory,
the sequence cannot be initiated, and an error is generated
There is a blocking SCPI query that can be used to check whether source list sequence was initiated
successfully (see "Remote Software Trigger (Remote command Only)" on page 2261)
Dependencies In Sequence Analyzer Mode, if Meas Setup, Include Source is set to YES, Source, List Sequencer,
Initiate Sequence is disabled
Not available in E7760B
Spectrum Analyzer Mode User's &Programmer's Reference 2237

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals