6 Input/Output
6.9 Calibration
Remote Command
:SYSTem:CALibration:ROW[1]|2|…|100:NAME <string>
:SYSTem:CALibration:ROW[1]|2|…|100:NAME?
Example
:SYST:CAL:ROW2:NAM "Monday AM Cal"
Notes 45 chars max; may not fit on display if max chars used
Dependencies The SCPI command is applied to the currently selected Cal Group. The subopcode is used to identify
the Cal row in the Cal Group
If the subopcode does not reference an existing Cal row in the Cal Group, the error message “-221,
Settings conflict; Subopcode does not reference an existing Cal row” is generated
Preset “Cal #”, where # is corresponding Cal number
State Saved Saved in instrument state
6.9.1.24 Last Cal
Displays the date and time the selected Cal was last executed. Read only field.
Remote Command
:SYSTem:CALibration:ROW[1]|2|…|100:LAST?
Example Return data and time Cal 2 was last executed:
:SYST:CAL:ROW2:LAST?
Notes Returns a string containing the date and time the Cal was executed. If the Cal has never been
executed, or any of the settings are changed, SCPI returns an empty string, and the front panel
displays “---"
Dependencies The SCPI query applies to the currently selected Cal Group
If the subopcode does not reference an existing Cal row in the Cal Group, the error message “-221,
Settings conflict; Subopcode does not reference an existing Cal row” is generated
6.9.1.25 Cal Applied
Displays the status of a Cal once it is applied. Is either Yes or No, depending on if the
Cal was successfully applied or not. See RCalStatus for more details. If it is not
being applied, the field shows “---". Read-only field.
Remote Command
:SYSTem:CALibration:ROW[1]|2|…|100:CAPPlied?
Example Return Cal Stats of Cal 2:
:SYST:CAL:ROW2:CAPP?
Notes Returns a string containing the date and time the Cal was executed. If the Cal has never been
executed, or any of the settings are changed, SCPI returns an empty string, and the front panel
displays “---"
Dependencies The SCPI query applies to the currently selected Cal Group
If the subopcode does not reference an existing Cal row in the Cal Group, the error message “-221,
Settings conflict; Subopcode does not reference an existing Cal row” is generated
Spectrum Analyzer Mode User's &Programmer's Reference 2495

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals