3 Spectrum Analyzer Mode
3.5 ACP Measurement
command is sent, a "Setting conflict" warning is generated
Couplings When "Res BW" on page 936 Mode is Auto, this value is exactly the same as Res BW. When you
change this value, Res BW Mode also changes to Man
Preset When "Max Num of Offsets" on page 1010 is 12, the preset value of Offset G ~ L is 100kHz
100 kHz, 100 kHz, 100k Hz, 100 kHz | 100 kHz,100 kHz, 100 kHz,100 kHz, 100 kHz, 100 kHz
State Saved Saved in instrument state
Min/Max 1 Hz/8 MHz
Auto Function
Remote
Command
[:SENSe]:ACPower:OFFSet[1]|2:INNer:LIST:BANDwidth:RESolution:AUTO ON | OFF |
1 | 0,…
[:SENSe]:ACPower:OFFSet[1]|2:INNer:LIST:BANDwidth:RESolution:AUTO?
Subopcode: 1 = BTS/Downlink (Default), 2 = MS/Uplink
Example
:ACP:OFFS2:INN:LIST:BAND:RES:AUTO 1,1,1,1,1,1
:ACP:OFFS2:INN:LIST:BAND:RES:AUTO?
Preset When "Max Num of Offsets" on page 1010 is 12, the preset value of Offset G ~ L is 1
1, 1, 1, 1, 1, 1
State Saved Yes
Video BW
Lets you change the instrument post-detection filter (VBW).
Remote Command
[:SENSe]:ACPower:OFFSet[1]|2:INNer:LIST:BANDwidth:VIDeo <freq>,…
[:SENSe]:ACPower:OFFSet[1]|2:INNer:LIST:BANDwidth:VIDeo?
Subopcode: 1 = BTS/Downlink (Default), 2 = MS/Uplink
Example
:ACP:OFFS2:INN:LIST:BAND:VID 5MHz,5MHz,5MHz,5MHz,5MHz,5MHz
:ACP:OFFS2:INN:LIST:BAND:VID?
Notes The values shown in this table reflect the conditions after Mode Preset
Dependencies When "Meas Method" on page 970 is RBW, FAST or Fast Power, this cell is grayed-out and disabled.
If the cell is pressed, an advisory message is generated. If the equivalent remote command is sent, a
"Setting conflict" warning is generated
Preset When "Max Num of Offsets" on page 1010 is 12, the preset value of Offset G ~ L is 1MHz
1 MHz, 1 MHz, 1 MHz, 1 MHz, 1 MHz, 1 MHz
State Saved Yes
Min/Max 1 Hz/50 MHz
Auto Function
Spectrum Analyzer Mode User's &Programmer's Reference 1003