3 Spectrum Analyzer Mode
3.13 Phase and Amplitude vs Time Measurement
selected at the time the command is sent. Note that the Center Freq function in the
Frequency menu on the front panel always applies to the currently selected input.
Remote Command
[:SENSe]:FREQuency:IQ:CENTer <freq>
[:SENSe]:FREQuency:IQ:CENTer?
Example
:FREQ:IQ:CENT 30 MHz
Notes This command is the same in all modes, but the parameter is Measurement Global. So the value is
independent in each mode and common across all the measurements in the mode
Dependencies I/Q Center Freq is not available in VXT Models M9410A/11A
State Saved Saved in instrument state
Min -40.049995 MHz
Max 40.049995 MHz
CF Step
Changes the step size for "Center Frequency" on page 2006, and start/stop
frequency functions. Once a step size has been selected and the Center Frequency
function is active, the step keys (and the UP|DOWN parameters for Center Frequency
from remote commands) change Center Frequency by the step-size value.
Remote Command
[:SENSe]:FREQuency:CENTer:STEP[:INCRement] <freq>
[:SENSe]:FREQuency:CENTer:STEP[:INCRement]?
[:SENSe]:FREQuency:CENTer:STEP:AUTO OFF | ON | 0 | 1
[:SENSe]:FREQuency:CENTer:STEP:AUTO?
Example
:FREQ:CENT:STEP 500 MHz
:FREQ:CENT UP
increases the current center frequency value by 500 MHz
:FREQ:CENT:STEP?
:FREQ:CENT:STEP:AUTO ON
:FREQ:CENT:STEP:AUTO?
Notes Preset and Max values are depending on Hardware Options (503, 507, 508, 513, 526)
Dependencies Span, RBW, Center frequency
If the electronic/soft attenuator is enabled, any attempt to change the value of the center frequency
>3.6 GHz by pressing the Up-arrow key, fails and results in an advisory message. If the equivalent
SCPI command is sent, this same message is generated as part of a “-221, Settings conflict” warning
This control does not appear in 5G NR Mode
Couplings When auto-coupled in a non-zero span, the center frequency step size is set to 10% of the span.
When auto-coupled in zero span, the center frequency step size is set to the equivalent -3 dB RBW
value
Spectrum Analyzer Mode User's &Programmer's Reference 2011

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals