EasyManuals Logo

Keysight x-series User Manual

Keysight x-series
2852 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #834 background imageLoading...
Page #834 background image
3 Spectrum Analyzer Mode
3.4 Occupied BW Measurement
:CHP:SWE:ACQ:TIME?
Dependencies Available only on non-sweeping hardware
Couplings Coupled to Span, RBW, and VBW when in the Auto state
When you manually set a value when in the Auto state, the state automatically changes to Man
Preset Automatically calculated
State Saved Saved in instrument state
Min 100 ns
Max 4.00 ks
Auto Function
Remote
Command
[:SENSe]:<meas>:SWEep:ACQuisition:TIME:AUTO OFF | ON | 0 | 1
[:SENSe]:<meas>:SWEep:ACQuisition:TIME:AUTO?
<meas> is the identifier for the current measurement; any one of CHPower-
|ACPower|OBWidth|MONitor
Example Channel Power measurement:
:CHP:SWE:ACQ:TIME:AUTO OFF
Preset
ON
Sweep/Measure
Lets you toggle between Continuous and Single sweep or measurement operation.
The single/continuous state is Meas Global, so the setting affects all measurements.
The front-panel key Single/Cont performs exactly the same function
See "More Information" on page 835
Remote
Command
:INITiate:CONTinuous OFF | ON | 0 | 1
:INITiate:CONTinuous?
Example Put instrument into Single measurement operation:
:INIT:CONT 0
:INIT:CONT OFF
Put instrument into Continuous measurement operation:
:INIT:CONT 1
:INIT:CONT ON
Preset
ON
Note that :SYST:PRES sets :INIT:CONT to ON, but *RST sets :INIT:CONT to OFF
State Saved Saved in instrument state
Annunciation The Single/Continuous icon in the Meas Bar changes depending on the setting:
834 Spectrum Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)