3 Spectrum Analyzer Mode
3.5 ACP Measurement
And on the control it looks like this:
NOTE
If a Transducer Unit is set, it is displayed as Xducer Unit in the Y Axis Unit
dropdown. However, you can only
change
the Transducer Unit via the Edit
Correction dialog in the Input/Output, Corrections menu. In that dialog, tap
Settings then Transducer Unit. You can also turn off Transducer Unit from the
same menu, by selecting None.
If a Transducer Unit is set, it is displayed as Xducer Unit in the Y Axis Unit
dropdown. However, you can only
change
the Transducer Unit via the Edit
Correction dialog in the Input/Output, Corrections menu. In that dialog, tap
Settings then Transducer Unit. You can also turn off Transducer Unit from the
same menu, by selecting None.
The Transducer Units are:
Units Example
dBmV/m :UNIT:POW DBUVM
dBmA/m :UNIT:POW DBUAM
dBpT :UNIT:POW DBPT
dBG :UNIT:POW DBG
dBmA :UNIT:POW DBUA
None n/a
Ref Level Offset
Adds an offset value to the displayed reference level. The reference level is the
absolute amplitude represented by the top graticule line on the display.
894 Spectrum Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals