3 Spectrum Analyzer Mode
3.8 Spurious Emissions Measurement
Notes Comma-separated list of values
Preset
1, 1, 1, 1, 1, 1, 1, 1, 1, 1, 1, 1, 1, 1, 1, 1, 1, 1, 1, 1
State Saved Yes
Max 1000
Min/Max 1
Backwards
Compatibility
SCPI
[:SENSe]:SPURious[:RANGe][:LIST]:BWIDth:IMULti
Video BW
Sets the Video BW mode of the instrument. This can be Auto, where the instrument
determines the optimum setting, or Manual, where you determine the setting. This
parameter can send up to 20 values. The location in the list sent corresponds to the
range the value is associated with. When sending the remote command, missing
values are not permitted; that is, if you want to change values 2 and 6, then you
must send all values up to 6. Subsequent values will remain as they were.
The query always returns 20 values.
Remote
Command
[:SENSe]:SPURious[:RANGe][:LIST]:BANDwidth:VIDeo <freq>, <freq>, <freq>,
<freq>, <freq>, <freq>, <freq>, <freq>, <freq>, <freq>, <freq>, <freq>,
<freq>, <freq>, <freq>, <freq>, <freq>, <freq>, <freq>, <freq>
[:SENSe]:SPURious[:RANGe][:LIST]:BANDwidth:VIDeo?
Example
:SPUR:BAND:VID 1kHz, 10kHz, 100kHz, 1MHz, 1MHz, 1MHz, 1MHz, 3MHz, 3MHz,
3MHz, 3MHz, 3MHz, 3MHz, 3MHz, 3MHz, 3MHz, 3MHz, 3MHz, 3MHz, 3MHz
:SPUR:BAND:VID?
Preset SA, WCDMA, WLAN Modes: Automatically calculated
LTE, LTEAFDD, MSR Modes: 4.7kHz, 47kHz, 470kHz, 5MHz, 470kHz, 5MHz, 5MHz, 300kHz, 300kHz,
300kHz, 300kHz, 300kHz, 300kHz, 300kHz, 300kHz, 300kHz, 300kHz, 300kHz, 300kHz, 300kHz
LTETDD, LTEATDD Modes:
4.7kHz, 47kHz, 470kHz, 5MHz, 5MHz, 5MHz, 5MHz, 300kHz, 300kHz, 300kHz, 300kHz, 300kHz,
300kHz, 300kHz, 300kHz, 300kHz, 300kHz, 300kHz, 300kHz, 300kHz
5G NR Mode:
100 Hz, 1 kHz, 10 kHz, 100 kHz, 100 kHz, 100 kHz, 100 kHz, 300 kHz, 300 kHz, 300 kHz, 300 kHz, 300
kHz, 300 kHz, 300 kHz, 300 kHz, 300 kHz, 300 kHz, 300 kHz, 300 kHz, 300 kHz
State Saved Saved in instrument state
Min/Max 1 Hz/50 MHz
Backwards
Compatibility
SCPI
[:SENSe]:SPURious[:RANGe][:LIST]:BWIDth:VIDeo
Auto Function
1448 Spectrum Analyzer Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)