3 Spectrum Analyzer Mode
3.9 SEM Measurement
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:TEST?
Example
:SEM:OFFS:INN:LIST:TEST ABS, ABS, ABS, ABS, ABS, ABS
:SEM:OFFS:INN:LIST:TEST?
Notes Comma-separated list of values
Preset
ABS, ABS, ABS, ABS, ABS, ABS, ABS, ABS, ABS, ABS, ABS, ABS
State Saved Saved in instrument state
Range Absolute | Relative | Abs AND Rel | Abs OR Rel
Show Abs2 Limit
Same as "Show Abs2 Limit" on page 1683 under Limits.
Abs2 Start
Sets the 2nd absolute power level limit at the start frequency for the selected inner
offset, ranging from –200 to +50 dBm.
The fail condition for each inner offset channel is set remotely using:
[:SENSe]:SEMask:OFFSet[n]:INNer:LIST:TEST:SABSolute
You can turn off (not use) specific inner offset channels remotely using:
[:SENSe]:SEMask:OFFSet[n]:INNer:LIST:STATe
The query returns values currently set to the 2nd absolute power test limits.
When sending the command, missing values are not permitted; that is, if you want
to change values 2 and 6, you must send all values up to 6. Subsequent values
remain unchanged. The query always returns 12 values.
Remote
Command
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:STARt:SABSolute <real>,
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:STARt:SABSolute?
Example
:SEM:OFFS:INN:LIST:STAR:SABS -12.50 dBm, -12.50 dBm, -24.50 dBm, -11.50 dBm,
-11.50 dBm, -11.50 dBm
:SEM:OFFS:INN:LIST:STAR:SABS?
Notes Comma-separated list of values
OFFSet 1 is for BTS, 2 for MS. Default is BTS
Preset 0dBm, 0 dBm, 0 dBm, 0 dBm, 0 dBm, 0 dBm, 0 dBm, 0 dBm, 0 dBm, 0dBm, 0 dBm, 0 dBm|0 dBm, 0
dBm, 0 dBm, 0 dBm, 0 dBm, 0 dBm, 0dBm, 0 dBm, 0 dBm, 0 dBm, 0 dBm, 0 dBm
State Saved Saved in instrument state
Min -200 dBm
Max 50 dBm
Spectrum Analyzer Mode User's &Programmer's Reference 1683

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals