3 Spectrum Analyzer Mode
3.13 Phase and Amplitude vs Time Measurement
State Saved Saved in instrument state
Min -100 MHz
Max 100 MHz
Status Bits/OPC depend-
encies
Non-overlapped
Frequency Error Measurement Time
Sets the acquisition time for "Measure Frequency Error" on page 2015.
Remote Command
[:SENSe]:PAVTime:FERRor:TIME <time>
[:SENSe]:PAVTime:FERRor:TIME?
Example
:PAVT:FERR:TIME 1 us
:PAVT:FERR:TIME?
Preset 10 ms
State Saved Saved in instrument state
Min/Max 100 ns/500 ms
Segment Frequency Error Correction
Sets the Segment Frequency Error Correction to On or Off. When ON, frequency
error is calculated simultaneously with each sample point’s IQ values. The phase
results are compensated with this real time frequency error, as well as the
Frequency Error entered by user or calculated by "Measure Frequency Error" on
page 2015.
Remote
Command
[:SENSe]:PAVTime:FERRor:CORRection[:STATe] OFF | ON | 0 | 1
[:SENSe]:PAVTime:FERRor:CORRection[:STATe]?
Example
:PAVT:FERR:CORR ON
:PAVT:FERR:CORR?
Preset
OFF
State Saved Saved in instrument state
Sync Type (Models with Multiple Receivers only)
Sets the Sync Type when multiple receivers in one instrument need to sync at data
acquisition. This setting is available when the instrument supports multiple
receivers. Only one receiver can be set as Primary, and the others can only be set as
Secondary.
2016 Spectrum Analyzer Mode User's &Programmer's Reference